Laminar Wash MINI1000

Curiox Biosystems
Cell
Processing
SignAC
Med. Bldg. B

Overview

Automatic cell washing system with no need of centrifugation. Laminar flow generated by double nozzles allows mild cell washing. Up to 8 samples can be processed at one time.

Features

  • Reduces cell damage and cell loss
  • A small number of cells can be washed (100 cells/well)
  • Short processing time (6 minutes for 16 samples)
  • High cleaning capacity
  • Improved single-cell sequencing data due to reduced background

How to use

  • Available to ASHBi PI group members. Reservations are required.
  • Please provide consumables (16-well strips, tips, and washing buffer) yourself.
  • Required solutions: 70% ethanol (1% Tween20), water (1% Tween20), and cell washing buffer. Please prepare them yourself.

Usage rules

  • Please leave usage logs.
  • When using 70% ethanol (containing 1% Tween 20), water (containing 1% Tween 20), and cell washing buffer, be sure to treat them with a 0.22 µm filter before use. (To prevent clogging of equipment due to debris)
  • When you prepare washing buffers in reagent bottles that can be attached to the buffer inlet bottle cap (GL45 standard screw-top bottles), you may reuse the cap.
  • If the system has not been used for more than one month, priming with 70% ethanol (1% Tween 20) should be performed twice instead of once at the startup.
  • When you apply cells with strong adhesion or cells containing a large amount of debris, please wash them with PURELOX diluted 12 times with MilliQ (SignAC will prepare) at the shutdown.
  • Wash reagent bottles (including buffer inlet bottle caps and tubes) with MilliQ after use.
  • Dispose of waste liquid after use.